Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the wordpress-seo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /var/www/html/wp-includes/functions.php on line 6260 Cycle Parameters For Decontaminating A Biological Safety Cabinet Using H2O2 Vapor | Contract Pharma
White Papers / Tech Papers

Cycle Parameters For Decontaminating A Biological Safety Cabinet Using H2O2 Vapor

Acumen White Papers - Volume 1 No. 2

Several studies have shown that hydrogen peroxide vapor (H2O2) can be useful in decontaminating HEPA filters, isolation chambers and centrifuge enclosures.1-3 However, before hydrogen peroxide can be used reliably in a biological safety cabinet (BSC), it is essential to establish the cycle parameters which allow full decontamination, and which minimize overall cycle time. This paper describes research which established the appropriate physical modifications and decontamination cycle parameters for the Baker Model SG-600, which is a Class II, Type A/B3 biological safety cabinet.